Table 2 Levels of VCA-IgA antibodies detected in the VCA-IgA antibody reference serum by the 6 commercial kits

From: Preliminary evaluation of a candidate international reference for Epstein–Barr virus capsid antigen immunoglobulin A in China

Manufacturer

 

Cut-off value

Reference serum

Ratio†

Within-run CV

Total CV

 

(arbitrary units)*

(arbitrary units)

High

Low

High

Low

Euroimmun

ELISA

1.1

13.0

11.8

3.98%

6.11%

6.24%

6.52%

Beier**

ELISA

1.1

3.2

2.9

5.03%

19.66%

13.77%

43.11%

Tarcine

ELISA

0.15

1.4

9.3

4.95%

7.58%

8.27%

7.22%

Antu**

ELISA

0.15

3.5

23.3

4.62%

6.45%

13.74%

16.85%

New Industries

CLIA

4

12.0

3.0

2.47%

4.34%

5.95%

9.86%

YHLO

CLIA

1.1

5.6

5.1

4.76%

5.22%

3.98%

4.87%

  1. *VCA-IgA antibody levels detected in the reference serum at the highest concentration tested in each kit
  2. †Ratio was calculated as reference serum arbitrary units divided by manufacturer cutoff value for each kit
  3. **Precision was calculated by OD value of each test when all the negative and positive controls met manufacturer requirements