Skip to main content

Table 4 Stability of samples diluted in media A and B

From: Validation of Novaprep® HQ+ liquid-based cytology medium for high-risk human papillomavirus detection by hc2

 

RLU/CO

 

18 °C to 25 °C

2 °C to 8 °C

40 °C

-15 °C to -30 °C

Weeks

A1

A2

A3

B1

B2

B3

A1

A2

A3

B1

B2

B3

A1

A2

A3

B1

B2

B3

A1

A2

A3

B1

B2

B3

0

41

42

36

37

37

39

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

2

51

48

45

36

47

62

49

52

48

36

41

41

51

67

51

48

59

48

52

46

46

48

50

48

4

51

42

47

44

46

53

54

49

44

36

45

43

53

54

50

49

39

43

47

45

48

49

57

50

6

49

51

46

46

46

50

53

52

52

31

35

35

47

44

52

45

49

45

-

-

-

-

-

-

8

51

54

50

41

50

43

48

55

40

35

38

32

54

49

55

40

46

52

-

-

-

-

-

-

10

41

41

42

38

36

38

41

45

39

28

36

33

42

40

42

40

42

38

-

-

-

-

-

-

12

45

42

43

35

40

47

49

40

46

29

33

29

50

47

47

37

43

42

-

-

-

-

-

-

16

48

52

50

42

43

33

49

51

47

34

31

36

49

48

42

44

47

42

-

-

-

-

-

-

24

34

33

33

33

29

31

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

28

44

43

42

43

39

36

-

-

-

-

-

-

-

-

-

-

-

 

43

45

42

40

40

43

  1. RLU/CO relative light units/cutoff